Hanger-Bar 2P 45cm
Check against the official notice below — it lists the full affected range.
Human Care USA, Inc. is recalling the Hanger-Bar 2P 45cm due to a manufacturing defect in which Loctite adhesive was not applied to the set screw. This defect can cause the set screw to loosen, creating a potential fall hazard. The recall affects 10 devices with Part Number 72754, UPC Code 881608800397, EAN Code 7331599999707, and Lot Number 33218501 that were distributed nationwide in California, Illinois, Louisiana, Massachusetts, Nevada, Ohio, Oklahoma, Pennsylvania, Texas, Utah, Virginia, and West Virginia.
The source document does not specify a remedy or what affected consumers should do regarding this recall.
Written by software from the official notice below. Not reviewed by a lawyer.
The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard.
Reason for recall: The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard. Distribution: U.S. Nationwide distribution in the states of CA, IL, LA, MA, NV, OH, OK, PA, TX, UT, VA, and WV. Quantity: 10 devices Lot/code info: Part Number 72754 UPC Code: 881608800397 EAN Code: 7331599999707 Lot Number: 33218501
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.