← Search recalls
FDA (Device enforcement)Recalled 2023-11-28class ii

XTANT Medical Holdings, Inc: Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.

Hazard

Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.

Details from the source

Reason for recall: Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging. Distribution: US distribution in MI. Quantity: 6 Lot/code info: UDI-DI: 00810076564005, Lot: AP23000143

UPC codes

00810076564005

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.