← Search recalls
FDA (Device enforcement)Recalled 2019-01-17closedclass ii

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF): Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer

Is my product affected?

Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer

  • UPC 00643169121973
  • UPC 0643169166974
  • UPC 00643169166981
  • UPC 00643169166998
  • UPC 00643169167001
  • UPC 00643169167018
  • UPC 00643169167025
  • UPC 00643169167049
  • UPC 00643169167063
  • UPC 00643169167070
  • UPC 00643169167087
  • UPC 00643169167094
  • UPC 00643169167100
  • UPC 00643169225732
  • UPC 00643169225749
  • UPC 00643169225756
  • UPC 00643169225763
  • UPC 00643169225770
  • UPC 00643169369030

Check against the official notice below — it lists the full affected range.

What happened

Medtronic is recalling certain Model 26907 power supply cable plugs used with the Model 29901 CareLink Encore programmer. An electrical short circuit can develop in the low voltage direct current connector, causing it to heat up, generate smoke, melt, and potentially disconnect from the programmer. A total of 731 units have been distributed across the United States and multiple international locations.

The affected power supply cables are identified by specific lot and serial numbers beginning with "JVD" and corresponding Universal Product Numbers (UPNs) listed in the recall documentation. The recall record does not specify a remedy for consumers to pursue.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

There is a potential for an electrical short circuit to develop in a subset of 479 Model 26907 power supply cable plugs that connect to the Medtronic Model 29901 CareLink Encore programmer. A failure mode of the power supply has been found where an electrical short occurs on the low voltage direct current connector and results in the connector heating up, generating smoke, melting and in some cases disconnecting from the programmer.

Details from the source

Reason for recall: There is a potential for an electrical short circuit to develop in a subset of 479 Model 26907 power supply cable plugs that connect to the Medtronic Model 29901 CareLink Encore programmer. A failure mode of the power supply has been found where an electrical short occurs on the low voltage direct current connector and results in the connector heating up, generating smoke, melting and in some cases disconnecting from the programmer. Distribution: US and Armenia, Bulgaria, Croatia, Georgia, Hungary, India, Iran, Jordan, Kazakhstan, Latvia, Lebanon, Oman, Pakistan, Qatar, Romania, Saudi Arabia, Syrian Arab Republic Quantity: 731 units Lot/code info: UPN: 00643169121973 Lot/Serial Numbers: JVD000376P JVD000415P JVD000242P JVD000262P JVD000264P JVD000272P JVD000278P JVD000283P JVD000289P JVD000291P JVD000293P JVD000295P JVD000300P JVD000311P JVD000316P JVD000320P JVD000330P JVD000332P JVD000333P JVD000334P JVD000337P JVD000338P JVD000344P JVD000346P JVD000348P JVD000350P JVD000351P JVD000352P JVD000354P JVD000356P JVD000357P JVD000359P JVD000360P JVD000361P JVD000362P JVD000363P JVD000364P JVD000367P JVD000369P JVD000373P JVD000377P JVD000379P JVD000381P JVD000382P JVD000383P JVD000384P JVD000388P JVD000390P JVD000392P JVD000394P JVD000407P JVD000408P JVD000409P JVD000410P JVD000411P JVD000412P JVD000413P JVD000417P JVD000419P JVD000420P JVD000421P UPN: 0643169166974 Lot/Serial Numbers: JVD000331P JVD000426P JVD000430P UPN: 00643169166981 Lot/Serial Numbers: JVD000273P JVD000275P JVD000299P JVD000307P JVD000339P UPN: 00643169166998 Lot/Serial Numbers: JVD000245P JVD000265P JVD000297P JVD000301P JVD000313P JVD000387P JVD000391P JVD001090P JVD001137P UPN: 00643169167001 Lot/Serial Numbers: JVD000353P JVD000395P UPN: 00643169167018 Lot/Serial Numbers: JVD000386P JVD000389P JVD000425P JVD000432P JVD000269P JVD000290P JVD000326P UPN: 00643169167025 Lot/Serial Numbers: JVD000439P JVD000308P JVD000385P JVD001089P JVD000370P UPN: 00643169167049 Lot/Serial Numbers: JVD000322P JVD000378P JVD000418P JVD000434P UPN: 00643169167063 Lot/Serial Numbers: JVD000298P JVD000305P JVD000318P JVD000342P JVD000355P UPN: 00643169167070 Lot/Serial

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.