← Search recalls
FDA (Device enforcement)Recalled 2012-09-28closedclass ii

Ventana Medical Systems Inc: VENTANA (a member of the Roche Group) BenchMark ULTRA; REF/MU 750-600; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is i

Is my product affected?

VENTANA (a member of the Roche Group) BenchMark ULTRA; REF/MU 750-600; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or

  • UPC 05342716001

Check against the official notice below — it lists the full affected range.

What happened

Ventana Medical Systems Inc. is recalling 1,735 VENTANA BenchMark ULTRA and DISCOVERY ULTRA automated staining platforms distributed worldwide in the USA and Germany. These research-use-only instruments have a potential defect where waste fluid can overflow during decontamination cycles. This overflow could cause fluid to migrate across internal tubing and wiring, potentially contacting electrical circuitry and creating electrical shorting conditions.

The affected instruments are identified by catalog number 750-600 and part number 05342716001, with specific serial numbers listed in the recall documentation. The recall source does not specify a remedy for affected consumers.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Potiential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform. This type of overflow event can lead to migration of fluid across tubing and wiring within the instrument cabinet resulting in contact with electrical circuitry. This may create shorting conditions and u

Details from the source

Reason for recall: Potiential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform. This type of overflow event can lead to migration of fluid across tubing and wiring within the instrument cabinet resulting in contact with electrical circuitry. This may create shorting conditions and u Distribution: Worldwide distribution: USA (nationwide) and country of: Germany. Quantity: 1735 instruments Lot/code info: Cat#750-600; PN 05342716001; Serial #'s: 310724, 310735, 311187, 312055, 310373, 311052, 311171, 311953, 310750, 311266, 311267, 311564, 311565, 311596, 311304, 310481, 311924, 310607, 311627, 310527, 311436, 311742, 310132, 311432, 310034, 310408, 310687, 310697, 311343, 311958, 310728, 310778, 311991, 312070, 310471, 311326, 311952, 310103, 310777, 311446, 310924, 311303, 311528, 310490, 310913, 311529, 311494, 310139, 310281, 310465, 310275, 310617, 310665, 310716, 311853, 311871, 312137, 312101, 310240, 310237,310238, 311571, 311404, 310356, 310023, 312164, 312165, 310065, 310231, 310711, 310706, 310707, 310703, 311509, 310330, 310633, 311811, 311810, 310977, 311194, 311198, 310208, 310732, 310740, 310187, 311462, 311935, 310145, 311037, 310346, 310585, 310694, 310699, 310701, 310702, 310705, 310709, 310710, 310704, 311106, 311110, 311111, 311419, 311484, 311527, 310345, 311381, 311353, 310480, 311869, 311921, 311458, 310315, 310412, 311988, 310618, 310397, 310399, 311868, 311919, 312139, 312141, 312142, 311270, 310069, 311466, 311929, 310083, 310094, 310154, 311904, 310588, 312098, 312100, 310814, 311731, 311783, 311785, 312096, 312104, 312153, 310227, 310781, 310174, 311403, 311080, 311081, 311121, 311122, 311169, 311188, 311189, 311190, 311242, 311245, 311298, 311926, 310403, 310718, 311416, 311417, 312105, 312112, 310796, 310832, 310824, 311135, 311248, 312066, 310895, 310267, 310272, 311481, 311771, 311772, 310882, 311186, 311937, 310344, 311439, 311505, 310146, 3105

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.