Isopure Corp: Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifugal Pump Model #110192-3

Is my product affected?

Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifugal Pump Model #110192-3

  • UPC 00852456007092

Check against the official notice below — it lists the full affected range.

What happened

Isopure Corp is recalling 15 Acidified Concentrate Distribution and Storage System (ACDS) units equipped with Finish Thompson Centrifugal Pump Model #110192-3 that were distributed across 13 U.S. states. The recall addresses a manufacturing defect where the magnet ring may not be evenly encapsulated in plastic during molding, potentially causing partial axial displacement. This uneven encapsulation could lead to uneven wear during operation, expose the magnet to the solution, and result in corrosion that introduces iron into the system.

The affected units have specific serial numbers ranging from AC-837874 through AC-838746, each paired with corresponding pump serial numbers. The systems were distributed to Alabama, Florida, Georgia, Massachusetts, Maryland, Minnesota, North Carolina, New Mexico, New York, Ohio, Oklahoma, Texas, and Virginia. The source document does not specify a remedy for affected consumers.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.

Details from the source

Reason for recall: There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron. Distribution: The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA. Quantity: 15 Lot/code info: ACDS UDI: (01)00852456007092; ACDS Serial Number AC-837874 with Pump SN 178745C18; ACDS Serial Number AC-834638 with Pump SN 180212D18; ACDS Serial Number AC-837654 with Pump SN 178324B18; ACDS Serial Number AC-837864 with Pump SN 178743C18; ACDS Serial Number AC-837664 with Pump SN 178322B18; ACDS Serial Number AC-838050 with Pump Serial Number 178752C18; ACDS Serial Number AC-838030 with Pump SN 178746C18; ACDS Serial Number AC-838020 with Pump SN 178748C18; ACDS Serial Number AC-838173 with Pump SN 178744C18; ACDS Serial Number AC-838143 with Pump SN 178325B18; ACDS Serial Number AC-838260 with Pump SN 180211D18; ACDS Serial Number AC-838740 with Pump SN 180209D18; ACDS Serial Number AC-838743 with Pump SN 180207D18; ACDS Serial Number AC-838749 with Pump SN 180215D18; ACDS Serial Number AC-838746 with Pump SN 180206D18

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify