← Search recalls
FDA (Device enforcement)Recalled 2023-12-29class iii

Philips North America: Ingenia Ambition X, Model No. 782109

Hazard

An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.

Details from the source

Reason for recall: An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel. Distribution: Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia. Quantity: 7 US; 16 ROW total Lot/code info: Model No. 782109 UDI-DI 00884838098350 Serial Number 47785

UPC codes

00884838098350

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify