An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.
Reason for recall: An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel. Distribution: Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia. Quantity: 7 US; 16 ROW total Lot/code info: Model No. 781178 UDI-DI N/A Serial Number 20150; Model No. 781196 UDI-DI N/A Serial Nos. 22270 22095 21113; Model No. 781296 UDI-DI 00884838004108 Serial Nos. 33406 33547 32773 32789 32388
00884838004108
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.