FDA (Device enforcement)Recalled 2023-12-19class ii

Beckman Coulter, Inc.: iQ200 Series Urine Microscopy Analyzers, Part Numbers: a) 700-3320, b) 700-3370, c) 700-3345, d) 700-3347, e) C10683, f) 700-3375, g) 700-3325, h) C10684

Hazard

Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.

Details from the source

Reason for recall: Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results. Distribution: US and Australia, Brazil, Canada, China, France, Germany, Ghana, India, Israel, Italy, Lebanon, Lithuania, Mexico, Panama, Qatar, Romania, Saudi Arabia, Spain, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland, Quantity: 112 units Lot/code info: a) 700-3320, UDI/DI 10837461001188; b) 700-3370, UDI/DI 10837461001324; c) 700-3345, UDI/DI 10837461001300; d) 700-3347, UDI/DI 10837461001317; e) C10683, UDI/DI 15099590697129 f) 700-3375, UDI/DI 10837461001355; g) 700-3325, UDI/DI 10837461001256; h) C10684, UDI/DI 15099590695484; All serial numbers

UPC codes

10837461001188, 10837461001324, 10837461001300, 10837461001317, 15099590697129, 10837461001355, 10837461001256, 15099590695484

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify