← Search recalls
FDA (Device enforcement)Recalled 2024-12-02class ii

Moximed, Inc.: MISHA Knee System Implant Small, Left, REF: 2-1001. MISHA Knee System Implant Small, Right, REF: 2-1002.

Hazard

Knee implant may fracture due to a supplier-related issue with one component that has nonconforming geometry at the distal end of the absorber component that may result in a reduced wall thickness and microcracks, which may lead to new or worsening symptoms of discomfort, swelling / edema, pain or stiffness, which may necessitate removal of the device.

Details from the source

Reason for recall: Knee implant may fracture due to a supplier-related issue with one component that has nonconforming geometry at the distal end of the absorber component that may result in a reduced wall thickness and microcracks, which may lead to new or worsening symptoms of discomfort, swelling / edema, pain or stiffness, which may necessitate removal of the device. Distribution: US Nationwide distribution in the states of CA, OR, NY, OH, TX, CO, NY, VA. Quantity: 115 Lot/code info: REF/UDI-DI/Lot(Expiration): 2-1001/00856047005795/23032901(29-Mar-25), 23092102(21-Sep-25), 24032601(26-Mar-26). 2-1002/00856047005801/23032902(29-Mar-25), 23092503(25-Sep-25), 24040801(8-Apr-26).

UPC codes

00856047005795, 00856047005801

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify