← Search recalls
FDA (Device enforcement)Recalled 2023-12-21class ii

Cardinal Health 200, LLC: Cardinal Health Suction Canister Liner, MEDI-VAC CRD 1000 mL Suction Canister Liner Kit with 6 mm 1.8 m Tube, REF 65651-517

Hazard

The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.

Details from the source

Reason for recall: The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister. Distribution: US, Canada, Australia and New Zealand Quantity: 74515 Lot/code info: 50630140152670 (case), Lot Numbers: J308-625, J308-629, J308-637, J309-616, J309-634, J309-646, J309-647, J310-603, J310-618

UPC codes

50630140152670

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify