← Search recalls
FDA (Device enforcement)Recalled 2023-11-10class ii

Smiths Medical Asd Inc: 1050 STOPCOCK LEFT ROTATOR OFF HANDLE, List Number MX1431MRL

Hazard

There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.

Details from the source

Reason for recall: There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication. Distribution: Worldwide distribution - US Nationwide and the countries of Canada, UAE. Quantity: 31,685 total Lot/code info: Lot Number: 4387166

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify