← Search recalls
FDA (Device enforcement)Recalled 2021-12-10class ii

Ortho-Clinical Diagnostics, Inc.: VITROS Chemistry Products VersaTips, Catalog No. 6801715 A VersaTip sales unit is composed of 1000 tips per box.

Is my product affected?

VITROS Chemistry Products VersaTips, Catalog No. 6801715 A VersaTip sales unit is composed of 1000 tips per box.

  • UPC 10758750006786

Check against the official notice below — it lists the full affected range.

What happened

Ortho-Clinical Diagnostics is recalling VITROS Chemistry Products VersaTips (Catalog No. 6801715) due to shape deformities in one lot that may reduce the volume of sample or reagent dispensed by up to 16 percent. If the resulting measurement errors go undetected, patients could receive reduced drug dosages or receive false positive or negative results for bloodborne infectious disease tests, particularly when results are close to the cut-off value.

The affected lot (Lot 1039734, UDI 10758750006786) consists of 4,125 units and was distributed nationwide in the United States as well as to Canada. The source does not specify a remedy for affected consumers.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

The lot has a shape deformity which may lead to a 16% reduction in the volume of the sample or reagent dispensed. In cases where the error is not detected and the affected results are not suppressed, clinical impact may include reduced drug dosage, or false positive or false negative results for bloodborne infectious disease assays when sample results are close to the cut-off value.

Details from the source

Reason for recall: The lot has a shape deformity which may lead to a 16% reduction in the volume of the sample or reagent dispensed. In cases where the error is not detected and the affected results are not suppressed, clinical impact may include reduced drug dosage, or false positive or false negative results for bloodborne infectious disease assays when sample results are close to the cut-off value. Distribution: Domestic distribution nationwide. Foreign distribution to Canada. Quantity: 4125 units Lot/code info: UDI 10758750006786 Lot 1039734

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.