FDA (Device enforcement)Recalled 2021-12-30class ii

RAYSEARCH LABORATORIES AB: RayStation/RayPlan- designed for treatment planning and analysis of radiation therapy. Software Version: 6.0.0.24 , 6.1.0.26 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0

Is my product affected?

RayStation/RayPlan- designed for treatment planning and analysis of radiation therapy. Software Version: 6.0.0.24 , 6.1.0.26 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.61 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 . Including Service Paks: Product name (build

  • UPC 07350002010013
  • UPC 07350002010082
  • UPC 07350002010075
  • UPC 07350002010242
  • UPC 07350002010068
  • UPC 07350002010112
  • UPC 07350002010136
  • UPC 07350002010129
  • UPC 07350002010204
  • UPC 07350002010235
  • UPC 07350002010174
  • UPC 07350002010266
  • UPC 07350002010297

Check against the official notice below — it lists the full affected range.

What happened

RaySearch Laboratories AB is recalling RayStation/RayPlan software used for treatment planning and analysis of radiation therapy. The software versions affected include 6.0.0.24, 6.1.0.26, 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0, and several later versions through 9.2.0.483. A total of 75 licenses were distributed nationwide across the United States.

The software contains a defect where selecting a new primary image set while a cine loop is running causes that primary image set to be displayed as both the primary and secondary image set in all side-by-side views. The same problem occurs when opening a new patient or case while the cine loop is running. The recall record does not specify a remedy for affected consumers.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

If a new primary image set is selected while the cine loop is running, the primary image set will be displayed as both primary and secondary image set in all side-by-side views. This will also be true for any new patient or case opened while the cine loop is running

Details from the source

Reason for recall: If a new primary image set is selected while the cine loop is running, the primary image set will be displayed as both primary and secondary image set in all side-by-side views. This will also be true for any new patient or case opened while the cine loop is running Distribution: US Nationwide distribution. Quantity: 75 Licenses Lot/code info: Software: 6.0.0.24, 6.1.0.26, 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113, 9.1.0.933, 9.2.0.483 UDI: 07350002010013 07350002010082 07350002010075 07350002010242 07350002010068 07350002010112 07350002010136 07350002010129 07350002010204 07350002010235 07350002010174 07350002010266 07350002010297 07350002010013 07350002010082 07350002010075 07350002010242 07350002010068

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.