← Search recalls
FDA (Device enforcement)Recalled 2016-09-30closedclass ii

GF Health Products, Inc: Lumex, Patriot Full Electric Homecare Bed, & Patriot LX Full Electric Homecare Bed containing the Okin Dumat 3 Head/Foot Motors. Description Part Number Motor and Pendant Patr

Is my product affected?

Lumex, Patriot Full Electric Homecare Bed, & Patriot LX Full Electric Homecare Bed containing the Okin Dumat 3 Head/Foot Motors. Description Part Number Motor and Pendant Patriot Full Electric 120V 690-3001-943 / 690-7001-943 Boxed Full Electric 220V Motor 690-3220-943 Motor and Pendant Patriot LX F

  • UPC 697142145108
  • UPC 697142147927
  • UPC 697144140132
  • UPC 697144140172
  • UPC 69204141262
  • UPC 69204141737

Check against the official notice below — it lists the full affected range.

What happened

GF Health Products, Inc. and OKIN America have initiated a voluntary field correction for OKIN America "DUOMAT 3" Head/Foot motors used in Lumex Patriot Full Electric Homecare Beds (Models US0458/US0468) and Patriot LX Full Electric Homecare Beds (Model US6000). The motors contain contaminated material in the plastic housing that may crack, causing the motor to stop moving the head section as designed. While no injuries have been reported, if the head section falls, an injury is unlikely but possible.

Approximately 3,547 units are affected, including 3,063 beds and 484 replacement motors, distributed nationwide in the USA and internationally to Puerto Rico, Bahamas, Canada, Chile, El Salvador, and Santo Domingo. The affected beds have specific serial number ranges, and replacement motor kits were shipped between May 4, 2015 and June 30, 2016. The record does not specify what remedy an affected consumer should pursue.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

GF Health Products, Inc., in conjunction with OKIN America, has initiated a voluntary field correction affecting the OKIN America "DUOMAT 3" Head/Foot motor used in the manufacture of our Lumex Patriot Full Electric Bed Model US0458/US0468 and Patriot LX Full Electric Bed Model US6000. The failure mode is directly related to contaminated material in the plastic housing used to produce the DUOMAT 3 motor. Contaminated motor housings may crack resulting in a motor that will no longer actuate the head section per the intended design. If the head section were to fall it is unlikely but possible an injury could occur. No injuries have been reported.

Details from the source

Reason for recall: GF Health Products, Inc., in conjunction with OKIN America, has initiated a voluntary field correction affecting the OKIN America "DUOMAT 3" Head/Foot motor used in the manufacture of our Lumex Patriot Full Electric Bed Model US0458/US0468 and Patriot LX Full Electric Bed Model US6000. The failure mode is directly related to contaminated material in the plastic housing used to produce the DUOMAT 3 motor. Contaminated motor housings may crack resulting in a motor that will no longer actuate the head section per the intended design. If the head section were to fall it is unlikely but possible an injury could occur. No injuries have been reported. Distribution: Worldwide Distribution - USA ( nationwide) Distribution including Puerto Rico and to the countries of : Bahamas, Canada, Chile, El Salvador and Santo Domingo. Quantity: 3,547 (3063 beds, 484 motors) Lot/code info: BED SERIAL NUMBER RANGE US0458 / US0468 Patriot Full Electric 120V 697142145108 through 697142147927 US0458 / US0468 Patriot Full Electric 220V 697144140132 through 697144140172 US6000 Patriot LX Full Electric 120V 69204141262 through 69204141737 SERVICE MOTOR KITS WITH SHIP DATES OF MAY 4, 2015 THROUGH JUNE 30, 2016 Description Part Number Motor and Pendant Patriot Full Electric 120V 690-3001-943 / 690-7001-943 Boxed Full Electric 220V Motor 690-3220-943 Motor and Pendant Patriot LX Full Electric 120V 690-8004-912

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.