During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.
Reason for recall: During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm. Distribution: US Nationwide distribution in the state of GA. Quantity: 6 devices Lot/code info: UDI/DI: (01) 07640164844589, Serial Numbers: 200221-040, 200221-041, 200221-042, 200221-043, 200221-044, 200221-045