← Search recalls
FDA (Device enforcement)Recalled 2025-09-25class ii

DJO, LLC: Cold Form Wrap: Back, REF: DP163CT03-BLK-L,XL, and Shoulder, REF: DP163CT02

Hazard

The IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at the seams and/or reach temperatures capable of causing a second degree burn.

Details from the source

Reason for recall: The IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at the seams and/or reach temperatures capable of causing a second degree burn. Distribution: US Nationwide distribution in the states of WI, TN, TX, MD, GA, UT, WA, MI, NC, NE, CA, MA, CO, SC, OH, IN, MN, NM, PA, NY, IL, FL, NJ, VA, AZ, VT, CT, OK. Quantity: 1,594 Lot/code info: REF/UDI-DI/Lot: DP163CT02/00888912164382/190523, 171221. DP163CT03-BLK-L,XL/00888912164481/150221, 180821

UPC codes

00888912164382, 00888912164481

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.