← Search recalls
FDA (Device enforcement)Recalled 2024-11-27class i

Trokamed GmbH: Sheath, 20 Fr., 160 mm (REF: WA2PS20L), gastroenterology-urology endoscopic access overtube.

Hazard

If the inflow speed is too quick, it could result in reduced outflow through the shaft due to blood, tissue fragments and kidney stone fragments, and quickly build up of intrarenal pressure.

Details from the source

Reason for recall: If the inflow speed is too quick, it could result in reduced outflow through the shaft due to blood, tissue fragments and kidney stone fragments, and quickly build up of intrarenal pressure. Distribution: US (Domestic) distribution in state of PA and OUS (foreign) to country of Germany. . Quantity: 42 units Lot/code info: REF: WA2PS20L; UDI-DI: 04251303810919; Lot Numbers: 154236

UPC codes

04251303810919

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify