← Search recalls
FDA (Device enforcement)Recalled 2023-12-05class ii

Ambu Inc.: Ambu aView 2 A Monitor, Catalog numbers 405011000 and 405011000US2

Hazard

Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.

Details from the source

Reason for recall: Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening. Distribution: Domestic distribution nationwide. Quantity: 11845 distributed US Lot/code info: Catalog No. 405011000 - UDI-DI 5707480145089 Catalog No. 405011000US2 - UDI-DI 5707480156757 All products distributed since launch (May 2020)

UPC codes

5707480145089, 5707480156757

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify