Micro-X Ltd.: MICRO-X Rover Mobile X-ray System, # MXU-RV19

Hazard

During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.

Details from the source

Reason for recall: During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam. Distribution: US Nationwide Distribution: TX, CA, OH, ID, WA, AZ Quantity: 23 Lot/code info: UDI-DI: 09357123000013. Serial Numbers: 262, 274, 320, 329, 237, 269, 271, 290, 292, 307, 310, 311, 312, 321, 322, 323, 327, 330, 331, 332, 333, 234, 266.

UPC codes

09357123000013

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify