Nobel Biocare Usa Llc: NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Is my product affected?

NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

  • UPC 07332747160529

Check against the official notice below — it lists the full affected range.

What happened

Nobel Biocare USA recalled 702 units of NobelReplace Conical Connection RP 4.3x11.5 mm implants (Catalog Number 36707) distributed worldwide to the U.S., Canada, Mexico, and Turkey. The recall was initiated due to 16 complaints that surgical drivers were not engaging properly with the implants. The company's investigation determined that an incorrect offset gauge may have been used during manufacturing of the affected implants.

This manufacturing issue could potentially result in surgical procedures being delayed or aborted. The affected implants are identified by lot numbers 13116495, 13116496, and 13116497. The source document does not specify what action affected consumers should take.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.

Details from the source

Reason for recall: Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure. Distribution: Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, MA, ME, MI, MN, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI. The countries of Canada, Mexico and Turkey. Quantity: 702 pieces Lot/code info: Catalog Number: 36707 UDI Code: (01)07332747160529(10)13116495(11)210720(17)260618; (01)07332747160529(10)13116496(11)210720(17)260618; (01)07332747160529(10)13116497(11)210720(17)260618; Lot Numbers: 13116495, 13116496, 13116497

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.