← Search recalls
FDA (Device enforcement)Recalled 2023-11-07class ii

Abiomed, Inc.: Abiomed Introducer Kit, 23 Fr, w/Hlx Vlv Sterile Individual, and Part of Pump Set (004334, 0046-0035, 0052-0002) Part Number: 0052-3021. Intended for introduction of the Impella Cathete

Hazard

Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis.

Details from the source

Reason for recall: Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis. Distribution: Nationwide Foreign: Country AT AU BE BR CA CA CH CZ DE DK ES FR FR HK IN IT MX MY NL NO PA SG TW UK Quantity: 193 single units; 1034 units within kits Lot/code info: UDI-DI: 00885672009786

UPC codes

00885672009786

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify