Multiple software issues which have the potential to lead to misdiagnosis due to image artifacts or incorrect image orientation labels, or the need for a CT rescan. These include: 1. Artifacts may occur on perfusion images, 2. Orientation label reversed in 2D images, 3. Auto voice does not work during the cardiac scan, 4. Auto voice and operators voice are too quiet, patient cant hear, 5. Not all of the images are reconstructed in some cases, 6. No image displayed on the console monitor after enabling CCT Single Scan, 7. Ring artifact appears in 32x0.625 collimation of bone scan when soft tissue view is selected, 8. A design defect where the error couch motion signal of couch firmware causes couch to identify current position as target position in next protocol, and 9. calcium scoring study split into separate series after loading to ISP.
Reason for recall: Multiple software issues which have the potential to lead to misdiagnosis due to image artifacts or incorrect image orientation labels, or the need for a CT rescan. These include: 1. Artifacts may occur on perfusion images, 2. Orientation label reversed in 2D images, 3. Auto voice does not work during the cardiac scan, 4. Auto voice and operators voice are too quiet, patient cant hear, 5. Not all of the images are reconstructed in some cases, 6. No image displayed on the console monitor after enabling CCT Single Scan, 7. Ring artifact appears in 32x0.625 collimation of bone scan when soft tissue view is selected, 8. A design defect where the error couch motion signal of couch firmware causes couch to identify current position as target position in next protocol, and 9. calcium scoring study split into separate series after loading to ISP. Distribution: Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, ME, MI, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, PA, RI, SC, TN, TX, UT, VT, WI, & Puerto Rico and the countries of Albania, Algeria, Argentina, Australia, Austria, Bhutan, Bolivia, Bosnia-Herz, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croa Quantity: 828 Lot/code info: UDI-DI: 00884838085015 (Model 728143) UDI-DI: 00884838105508 (Model 728144)
00884838085015, 00884838105508
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.