Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling.
Reason for recall: Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling. Distribution: Worldwide Quantity: 78532 units Lot/code info: GTIN 00613994483195, Serial Numbers: NGV502657H, NGV508302H, NGV612281H, NGV612282H, NGV717221H, NGV723564H, NGV727392H, NGV727397H, NGV727400H, NGV727906H, NGV730310H, NGV741180H; GTIN 00643169100824, Serial Numbers: NGV503518H, NGV503522H, NGV504577H, NGV504580H, NGV504581H, NGV507269H, NGV507271H, NGV507272H, NGV507273H, NGV507431H, NGV507433H, NGV507434H, NGV507436H, NGV508247H, NGV508248H, NGV508257H, NGV509489H, NGV509490H, NGV510082H, NGV512179H, NGV512180H, NGV502833H, NGV503294H, NGV502489H, NGV511809H, NGV510811H, NGV507369H, NGV511074H, NGV512393H, NGV502490H, NGV509057H, NGV506230H, NGV503295H, NGV507614H, NGV509098H, NGV509769H, NGV503297H, NGV502903H, NGV504121H, NGV511164H, NGV512019H, NGV502487H, NGV504523H, NGV503292H, NGV511885H, NGV511163H, NGV509373H, NGV502488H, NGV506232H, NGV510872H, NGV509770H, NGV510474H, NGV512471H, NGV508053H, NGV504241H, NGV507628H, NGV507635H, NGV503352H, NGV512033H, NGV512037H, NGV507723H, NGV508260H, NGV509172H, NGV509477H, NGV503596H, NGV504429H, NGV502450H, NGV507824H, NGV510692H, NGV505405H, NGV506725H, NGV511877H, NGV512666H, NGV505006H, NGV509513H, NGV500356H, NGV501018H, NGV503021H, NGV503099H, NGV503186H, NGV503187H, NGV503487H, NGV504710H, NGV504901H, NGV505165H, NGV508052H, NGV509450H, NGV509762H, NGV512254H, NGV513130H, NGV505328H, NGV506132H, NGV508199H, NGV504567H, NGV512052H, NGV501951H, NGV502809H, NGV502344H, NGV506202H, NGV505245H, NGV505772H, NGV508420H, NGV510444H, NGV510215H, NGV502774H, NGV504027H, NGV509803H
00613994483195, 00643169100824, 00643169345218, 00643169345225
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.