← Search recalls
FDA (Device enforcement)Recalled 2024-10-09class ii

Shimadzu Corp. Analytical & Measuring Instruments Div.: Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

Hazard

There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.

Details from the source

Reason for recall: There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing. Distribution: Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE. Quantity: 48 Lot/code info: Model/UDI-DI/Serial Numbers: SIL-40C XR CL/04540217420582/L2C316051002, L2C316051003, L2C316051004, L2C316051005, L2C316051006, L2C316051007, L2C316151012, L2C316151013, L2C316151014, L2C316151015. SIL-40C X3 CL/04540217420575/L2C305850001, L2C305950002, L2C306151025, L2C306051001, L2C306051002, L2C306051003, L2C306051004, L2C306051005, L2C306051007, L2C306051006, L2C306051008, L2C306051009, L2C306051010, L2C306051011, L2C306051012, L2C306051013, L2C306051014, L2C306051015, L2C306151017, L2C306151018, L2C306151022, L2C306151023, L2C306151024, L2C306151026, L2C306151028, L2C306151027, L2C306151029, L2C306151030, L2C306151031, L2C306151032, L2C306151033, L2C306151034, L2C306151035, L2C306151036, L2C306151037, L2C306151038, L2C306151040, L2C306151039. Firmware versions below 1.16.

UPC codes

04540217420582, 04540217420575

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify