← Search recalls
FDA (Device enforcement)Recalled 2018-09-26closedclass i

Medtronic Navigation, Inc.: StealthStation Cranial and Synergy Cranial Depth Gauge: StealthStation Cranial Optical Pocket Guide - Instructions for cranial procedures using optical localization Stealth

Is my product affected?

StealthStation Cranial and Synergy Cranial Depth Gauge: StealthStation Cranial Optical Pocket Guide - Instructions for cranial procedures using optical localization StealthStation Cranial version 3.0, 3.1 CE 0344 Rx Only 2018-02 9735489 Revision 3 Medtronic Navigation, Inc. Biopsy Needle Kit (973306

  • UPC 00613994408174
  • UPC 00613994546081
  • UPC 00643169038110
  • UPC 00613994467102
  • UPC 00643169037946
  • UPC 00613994467126
  • UPC 00613994792884
  • UPC 00643169510531
  • UPC 00613994944986
  • UPC 00643169192140
  • UPC 00643169332645
  • UPC 00643169510548
  • UPC 00643169353305
  • UPC 00643169503311
  • UPC 00643169481275
  • UPC 00643169502611
  • UPC 00643169641167
  • UPC 00643169513754
  • UPC 00643169510579
  • UPC 00643169641174
  • UPC 00643169490475
  • UPC 00643169490482
  • UPC 00643169505759
  • UPC 00643169505797
  • UPC 00643169505766
  • UPC 00643169505773
  • UPC 00643169505780
  • UPC 00643169481268
  • UPC 00643169510562
  • UPC 00643169510586
  • UPC 00643169510593
  • UPC 00643169510609
  • UPC 00643169510616
  • UPC 00885074238241
  • UPC 00613994231031
  • UPC 00643169625532
  • UPC 00643169625570
  • UPC 00643169777101
  • UPC 00643169625587
  • UPC 00643169625594

Check against the official notice below — it lists the full affected range.

What happened

Medtronic Navigation, Inc. is recalling approximately 5,487 StealthStation Cranial and Synergy Cranial systems due to an incorrect display of the Biopsy Depth Gauge in the software's Guidance View. This display error resulted in biopsies of healthy tissue being performed. The affected systems were distributed across the United States, U.S. territories, and multiple countries internationally, and involve specific software versions from 2.2.0 through 3.1.0 depending on the system model.

The recall document does not specify a remedy or recommended corrective action for affected consumers.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

An incorrect display of the Biopsy Depth Gauge graphical representation in the Guidance View using Cranial software, which resulted in the biopsy of healthy tissue.

Details from the source

Reason for recall: An incorrect display of the Biopsy Depth Gauge graphical representation in the Guidance View using Cranial software, which resulted in the biopsy of healthy tissue. Distribution: US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY OUS: Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bosnia, Brazil, Brunei Darussalam, Bulgaria, Canada, Chile, China, Colombia Quantity: 5487 impacted systems Lot/code info: Product Number/CFN Affected Software Versions 9733763 2.2.0, 2.2.5, 2.2.6, 2.2.7 Synergy Cranial 9735585 3.0.0, 3.0.1, 3.0.2, 3.1.0 StealthStation S7 Cranial Kit CFN and UDI: CFN GTIN Description 9733763 00613994408174 SOFTWARE, 9733763, SYNERGY CRANIAL S7 9734068 00613994546081 SFTWR KIT 9734068 SYN CRANI S7 PLNG STN 9734068 00643169038110 SFTWR KIT 9734068 SYN CRANI S7 PLNG STN 9733937 00613994467102 Software Kit 9733937 Synergy Cranial S7 9733937 00643169037946 Software Kit 9733937 Synergy Cranial S7 9733928 00613994467126 Upgrade Prgm 9733928 Treon to S7 220V 9734261 00613994792884 UPGRD PRGM 9734261 StealthStation S7 9734261-G02 00643169510531 UPGRD PRGM 9734261 StealthStation S7 9734849 00613994944986 UPGD PRGM 9734849 S7 CPTL ENH PRGM 9735064 00643169192140 UPGRD PRGM 9735064 KAISER TREON TO S7 9735328 00643169332645 UPGRD PRGM 9735328 Stealth S7 Touch¿ 9735328-G02 00643169510548 UPGRD PRGM 9735328 Stealth S7 Touch 9733937-G02 00643169353305 Software Kit 9733937 Synergy Cranial S7 9733763-G03 00643169503311 SOFTWARE, 9733763, SYNERGY CRANIAL S7 9733937-G03 00643169481275 SOFTWARE KIT 9733937 SYNERGY CRANIAL S7 9733928-G02 00643169502611 UPGRADE PRGM 9733928 TREON TO S7 220V 9734261-G03 00643169641167 UPGRD PRGM 9734261 StealthStation S7 9734849-G02 00643169513754 UPGD PRGM 9734849 S7 CPTL ENH PRGM 9735064-G02 00643169510579 UPGRD PRGM 9735064 KAISER TREON TO S7 9735328-G03 00643169641174 UPGRD PRGM 9735328 Stealth S7 Touch 9735065-G04 00643169490475 UPGRADE KIT 9735065

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify