← Search recalls
FDA (Device enforcement)Recalled 2024-10-18class ii

Zyno Medical LLC: Z-800 Infusion System. Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and bl

Hazard

Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.

Details from the source

Reason for recall: Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains. Distribution: This product is distributed only in the United States. Quantity: 34,994 units Lot/code info: Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF Lot Code: Lots: 201016280, 200515888, 200406416, 230920212, 200413400, 20040660, 200716968, 201016720, 20071632, 201211740, 201208256, 210319640, 210709900, 201211360, 2101221000, 210913560, 210913200, 231008352, 220205400, 220204548, 2211141000, 230206980, 230620800, 230420548, 230420136, 230525452, 230620200.

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.