Certain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while the device is running on battery power, the SimplyGo Mini will shut down and stop providing oxygen to the user. This may happen at a temperature within the specified operating temperature range of the SimplyGo Mini.
Reason for recall: Certain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while the device is running on battery power, the SimplyGo Mini will shut down and stop providing oxygen to the user. This may happen at a temperature within the specified operating temperature range of the SimplyGo Mini. Distribution: US Nationwide distribution in the states of GA, ID, KY, NJ, NY and the country of Canada. Quantity: 56 units (55 US, 1 OUS) Lot/code info: Product Code: 1116816; UDI-DI: 00606959032385; Serial Numbers: 2438V115830, 2438V115960, 2438V116033, 2438V115899, 2438V115832, 2438V115964, 2438V116034, 2438V115837, 2438V115965, 2438V116040, 2438V115840, 2438V115967, 2438V116044, 2438V115842, 2438V115968, 2438V116045, 2438V115844, 2438V115970, 2438V116046, 2438V115878, 2438V115972, 2438V116047, 2438V115890, 2438V115973, 2438V116048, 2438V115894, 2438V115974, 2438V116049, 2438V115895, 2438V115980, 2438V116050, 2438V115896, 2438V115981, 2438V116051, 2438V115897, 2438V115983, 2438V116052, 2438V115919, 2438V115984, 2438V116226, 2438V115937, 2438V115991, 2438V116229, 2438V115939, 2438V115992, 2438V116233, 2438V115942, 2438V115993, 2438V116249, 2438V115948, 2438V115994, 2438V116253, 2438V115957, 2438V115995, 2438V116258, 2438V117762;
00606959032385
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.