← Search recalls
FDA (Device enforcement)Recalled 2024-10-04class ii

Hobbs Medical, Inc.: Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4704

Hazard

Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

Details from the source

Reason for recall: Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle. Distribution: Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada. Quantity: 6 units Lot/code info: Catalog Number: 4704; UDI-DI: M84947040; Lot Numbers: H10-23-106, H11-23-059.

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify