FDA (Device enforcement)Recalled 2020-10-23class ii

Mako Surgical Corporation: Mako Integrated Cutting System (MICS) Handpiece, Catalog Number 209063, Software Version#: THA 3.0 and later, PKA 3.0 and later, and TKA 1.0 - Product Usage: The MICS Handpi

Is my product affected?

Mako Integrated Cutting System (MICS) Handpiece, Catalog Number 209063, Software Version#: THA 3.0 and later, PKA 3.0 and later, and TKA 1.0 - Product Usage: The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO (Mako robot) and to execute bone preparati

  • UPC 00848486030193

Check against the official notice below — it lists the full affected range.

What happened

Stryker has recalled the Mako Integrated Cutting System (MICS) Handpiece due to a characterization issue that may prevent proper verification of saw blade location before bone preparation during knee surgery procedures. This defect could cause the robotic surgical system to make bone cuts that differ from the surgeon's preoperative plan when using straight or angled saw attachments. The recall affects 367 units distributed worldwide, including throughout the United States and in Australia, Japan, Korea, the Netherlands, Chile, France, Germany, Luxembourg, and the UK.

The affected handpieces are identified by catalog number 209063 with software versions THA 3.0 and later, PKA 3.0 and later, and TKA 1.0, with specific serial numbers listed in the recall documentation. The source document does not specify any remedy or corrective action for consumers to take.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Handpiece. The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later). This issue may result in unsuccessful checkpoint verification when the Mako System attempts to verify the location of the saw blade prior to bone preparation. This issue may also cause a discrepancy during bone preparation in a Mako Total Knee or a Mako Partial Knee procedure in which the surgeon uses a straight or angled saw attachment with the MICS Handpiece. The surgeon may make cuts during bone preparation that differ from the surgeon s preoperative plan.

Details from the source

Reason for recall: Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Handpiece. The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later). This issue may result in unsuccessful checkpoint verification when the Mako System attempts to verify the location of the saw blade prior to bone preparation. This issue may also cause a discrepancy during bone preparation in a Mako Total Knee or a Mako Partial Knee procedure in which the surgeon uses a straight or angled saw attachment with the MICS Handpiece. The surgeon may make cuts during bone preparation that differ from the surgeon s preoperative plan. Distribution: Worldwide distribution - US Nationwide distribution and the countries of Australia, Japan, Korea, Netherlands, and Chile, and France, Germany Luxembourg, and UK Quantity: 367 units Lot/code info: (01) 00848486030193 (21)<serial number>, Serial numbers: 4200052, 4200095, 4200112, 4200117, 4200119, 4200167, 4200190, 4200193, 4200202, 4200203, 4200204, 4200226, 4200233, 4200235, 4200248, 4200250, 4200257, 4200269, 4200270, 4200288, 4200293, 4200305, 4200350, 4200373, 4200384, 4200387, 4200391, 4200397, 4200398, 4200400, 4200409, 4200436, 4200457, 4200459, 4200477, 4200488, 4200512, 4200540, 4200548, 4200550, 4200558, 4200624, 4200645, 4200653, 4200666, 4200717, 4200766, 4200774, 4200785, 4200789, 4200801, 4200806, 4200807, 4200816, 4200843, 4200856, 4200868, 4200904, 4200917, 4201000, 4201040, 4201068, 4201089, 4201102, 4201103, 4201121, 4201167, 4201218, 4201236, 4201238, 4201253, 4201296, 4201302, 4201366, 4201367, 4201376, 4201385, 4201408, 4201412, 4201442, 4201452, 4201454, 4201477, 4201490, 4201630, 4201661, 4201714, 4201752, 4201829, 4201831, 4201851, 4202009, 4202020, 4202038, 4202044, 4202062, 4202063, 4202122, 4202146, 4202162, 4202175, 4202315, 4202333, 4202626, 4202631, 4202659, 4202729, 4202742, 4202882, 4202987, 4203180, 4203198, 4203289, 4203424, 4203441, 4203649, 4203709, 4203728, 4203730, 4203731, 4203736, 4203807, 4203826, 4204015, 4204164, 4204389, 4204448, 4204539, 4204553, 4204577, 4204673, 4204812, 4204934, 4205093, 4205190, 4205380, 4205426, 4206219, 4206271, 4209023, 4209025, 4209044, 4209045, 4209046, 4209065, 4209066, 4209067, 4209068, 4209071, 4209149, 4209151, 4209155, 4209156, 4209157, 4209162, 4209164, 4209166, 4209167, 4209171, 4209185, 420

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify