FDA (Device enforcement)Recalled 2023-10-25class ii

Elekta, Inc.: ELEKTA AXESSE, REF XRT 2181, XRT 27891; Medical Charged-Particle Radiation Therapy System Accelerator, Linear

Hazard

Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.

Details from the source

Reason for recall: Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension. Distribution: worldwide Quantity: 47 units Lot/code info: UDI/DI N/A, Serial Numbers: 152140, 152617, 152691, 152004, 152683, 152994, 153935, 152620, 152809, 152854, 152867, 152885, 152964, 152974, 153192, 153217, 153527, 152499, 153668, 152519, 153118, 153102, 153204, 153166, 152017, 152112, 152325, 152749, 152754, 152755, 152756, 152860, 153006, 151874, 152538, 152839, 152921, 153149, 153267, 152736, 151793, 151802, 151812, 151960, 152600, 152678, 152838

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify