FDA (Device enforcement)Recalled 2023-10-25class ii

Elekta, Inc.: ELEKTA LIMITED Versa HD, REF XRT 2121, 2131; Medical Charged-Particle Radiation Therapy System Accelerator, Linear

Hazard

Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.

Details from the source

Reason for recall: Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension. Distribution: worldwide Quantity: 451 units Lot/code info: UDI/DI 05060191071529, Serial Numbers: 153396, 153671, 109003, 153383, 153384, 153055, 153980, 154001, 154002, 154003, 154004, 154272, 153104, 153188, 153243, 153745, 153836, 153872, 153876, 154123, 154148, 154199, 154250, 154260, 154267, 154269, 154449, 154498, 154550, 152951, 153140, 153525, 153526, 153532, 154218, 153621, 153951, 153949, 154028, 154548, 154239, 154262, 153413, 153829, 154163, 154101, 152777, 153037, 153142, 153144, 153177, 153186, 153187, 153189, 153277, 153364, 153399, 153425, 153426, 153471, 153510, 153537, 153578, 153625, 153661, 153752, 153762, 153782, 153797, 153808, 153816, 153847, 153858, 153950, 153986, 153997, 154046, 154070, 154084, 154085, 154127, 154178, 154202, 154217, 154280, 154372, 154374, 154381; UDI/DI 05060191071574, Serial Numbers: 153722, 109005, 109007, 153003, 153008, 153314, 153337, 153427, 153428, 153482, 153649, 153900, 154023, 154024, 154050, 154087, 154299, 154300, 153248, 153699, 154043, 154395, 154435, 154179, 154131, 153658, 153830, 154221, 154473, 153976, 153128, 153608, 154505, 154531, 154325, 153406, 153601, 153607, 153667, 153720, 153723, 153751, 153839, 153862, 153940, 153985, 154055, 154076, 154140, 154149, 154283, 154293, 154296, 154375, 154421, 154398, 152843, 152912, 153778, 153893, 153894, 154327, 154328, 154408, 153379, 153381, 152917, 153113, 153452, 153453, 154077, 154198, 154258, 154450, 153835, 153023, 153025, 153095, 153106, 153180, 153249, 153279, 153281, 153320, 153356, 153359, 153380, 153385, 153470, 153619

UPC codes

05060191071529, 05060191071574

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify