← Search recalls
FDA (Device enforcement)Recalled 2022-01-27closedclass ii

Integrity Implants Inc.: LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as n

Is my product affected?

LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. La

  • UPC 00818613028898
  • UPC 00818613028904
  • UPC 00818613028911
  • UPC 00818613028928
  • UPC 00818613028973
  • UPC 00818613028980
  • UPC 00818613028997
  • UPC 00818613029000
  • UPC 00818613029017
  • UPC 00818613026528
  • UPC 00818613026535
  • UPC 00818613026542
  • UPC 00818613026559
  • UPC 00818613026566
  • UPC 00818613026597
  • UPC 00818613026603
  • UPC 00818613026610
  • UPC 00818613026627
  • UPC 00818613026634
  • UPC 00818613026658
  • UPC 00818613026665
  • UPC 00818613026672
  • UPC 00818613026689
  • UPC 00818613026696

Check against the official notice below — it lists the full affected range.

What happened

Integrity Implants Inc. is recalling 1,369 LineSider Spinal System pedicle screws used in spinal fusion surgery. The screws, which range from 5.5mm to 6.5mm in diameter and various lengths, were distributed nationwide to hospitals and surgical centers in 15 states and Puerto Rico. The recall affects specific lot numbers manufactured between 2021 and 2022, identified by part numbers and UDI codes beginning with LS-N6MG.

The screws are being recalled because of a potential defect in the weld that joins the Upper Tulip Head component to the Lower Tulip Head component. If this weld separates, the Tulip Head could separate from the Screw Shank, potentially compromising the stability of the spinal fixation system. The recall record does not specify what action affected consumers should take in response to this recall.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.

Details from the source

Reason for recall: Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank. Distribution: Nationwide distribution to AZ, CA, CT, DC, FL, IL, IN, LA, MA, MN, NJ, OK, PR, TN, TX, VA. Quantity: 1,369 screws Lot/code info: a. 5.5mm x 35mm, Part Number: LS-N6MG5535, UDI: (01)00818613028898(10)02210093, lot Number: 2210093; UDI: (01)00818613028898(10)10200053, Lot Number: 10200053. b. 5.5mm x 40mm, Part Number: LS-N6MG5540, UDI: (01)00818613028904(10)02210094, lot Number: 2210094; UDI: (01)00818613028904(10)10200054, Lot Number: 10200054. c. 5.5mm x 45mm, Part Number: LS-N6MG5545, UDI: (01)00818613028911(10)02210095, lot Number: 2210095; UDI: (01)00818613028911(10)10200055, Lot Number: 10200055. d. 5.5mm x 50mm, Part Number: LS-N6MG5550, UDI: (01)00818613028928(10)02210096, lot Number: 2210096; UDI: (01)00818613028928(10)10200056, Lot Number: 10200056. e. 6.5mm x 35mm, Part Number: LS-N6MG6535, UDI: (01)00818613028973(10)02210097, lot Number: 2210097; UDI: (01)00818613028973(10)10200057, Lot Number: 10200057; UDI: (01)00818613028973(10)10200057-1, Lot Number: 10200057-1. f. 6.5mm x 40mm, Part Number: LS-N6MG6540, UDI: (01)00818613028980(10)02210098, lot Number: 2210098; UDI: (01)00818613028980(10)10200058, Lot Number: 10200058; UDI: (01)00818613028980(10)10200058-1, Lot Number: 10200058-1. g. 6.5mm x 45mm, Part Number: LS-N6MG6545, UDI: (01)00818613028997(10)02210099, lot Number: 2210099; UDI: (01)00818613028997(10)10200059, Lot Number: 10200059; UDI: (01)00818613028997(10)10200059-1, Lot Number: 10200059-1. h. 6.5mm x 50mm, Part Number: LS-N6MG6550, UDI: (01)008181613029000(10)02210100, lot Number: 2210100; UDI: (01)00818613029000(10)10200060, Lot Number: 10200060; UDI: (01)00818613029000(10)102

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify