FDA (Device enforcement)Recalled 2025-09-26class ii

Medistim Asa: Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System Model/Catalog Number: MQC04001, MQC14001, MQU14011, MQV12011 Software Version: Not applicable Product Description: U

Hazard

Printer Support Module (PSM) may be subject to insufficient soldering, which may cause loosening of the wires and system to shut down. Affected devices need to have their Printer Support Modules replaced.

Details from the source

Reason for recall: Printer Support Module (PSM) may be subject to insufficient soldering, which may cause loosening of the wires and system to shut down. Affected devices need to have their Printer Support Modules replaced. Distribution: Domestic distribution to CO, MI, MN, TN, TX. International distribution to Australia, Austria, Germany, Finland, India, Japan, Russia, China, Saudi Arabia. Quantity: 14 Lot/code info: Lot Code: Serial number; Model number 5042; MQC04001 5043; MQC04001 5044; MQC14001 5061; MQU14011 5062; MQU14011 5064; MQU14011 5065; MQU14011 5066; MQU14011 5067; MQC04001 5037; MQU14011 5073; MQV12011 5038; MQU14011 5045; MQU14011 5063; MQU14011

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.