PURE OXYGEN CONCENTRATOR, CHAD Drive, Models CH5000 and CH5000S
Check against the official notice below — it lists the full affected range.
Inovo, Inc. is recalling 13,140 units of its PURE OXYGEN CONCENTRATOR, CHAD Drive, Models CH5000 and CH5000S worldwide due to a hazard involving low oxygen purity. Post-market surveillance data shows that these devices have a complaint and service rate greater than 35% for producing oxygen output below the published specifications.
The affected devices include all serial numbers of Model CH5000 (UDI #00814470020013) and Model CH5000S (UDI #00814470020006). The source document does not specify a remedy or instructions for what affected consumers should do.
Written by software from the official notice below. Not reviewed by a lawyer.
Post-market surveillance for Pure Stationary Oxygen Concentrators model CHSOOO and CHSOOOS demonstrate an aggregate complaint/service rate greater than 35% for low oxygen purity (output oxygen below the published specification).
Reason for recall: Post-market surveillance for Pure Stationary Oxygen Concentrators model CHSOOO and CHSOOOS demonstrate an aggregate complaint/service rate greater than 35% for low oxygen purity (output oxygen below the published specification). Distribution: worldwide Quantity: 13,140 units Lot/code info: UDI #(s): 00814470020013 (model CHSOOO) and 00814470020006 (model CHSOOOS): All serial numbers
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.