← Search recalls
FDA (Device enforcement)Recalled 2024-09-04class ii

Bard Peripheral Vascular Inc: The EVSRF catheter is a sterile, single-use disposable medical device for endovenous radiofrequency ablation, a procedure in which a refluxing vein is thermally coagulate

Hazard

The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead to power being applied through the proximal 3/4th of the heating coil, bypassing the thermocouple, which will then cause the generator to overdrive the catheter in an attempt to reach desired temperature set point. The wire attachment error causes the device to reach temperatures exceeding the levels intended for therapeutic efficacy, while the console is falsely indicating that temperature has not yet been reached, thus increasing the risk of harm due to excessive temperature.

Details from the source

Reason for recall: The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead to power being applied through the proximal 3/4th of the heating coil, bypassing the thermocouple, which will then cause the generator to overdrive the catheter in an attempt to reach desired temperature set point. The wire attachment error causes the device to reach temperatures exceeding the levels intended for therapeutic efficacy, while the console is falsely indicating that temperature has not yet been reached, thus increasing the risk of harm due to excessive temperature. Distribution: Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI and WV. The countries of Australia, Singapore, Austria, Belgium, Czechia, Denmark, France, Germany, Greece, Ireland, Italy, Poland, Slovenia Quantity: 222,373 catheters Lot/code info: Catalog Number: VC-10A2.5-6F-100 UDI does: (01)00858254006015(17)241009(10)65497423 (01)00858254006015(17)241006(10)66030181 (01)00858254006015(17)241011(10)66599182 (01)00858254006015(17)241017(10)66634308 (01)00858254006015(17)241020(10)66784768 (01)00858254006015(17)241026(10)66793709 (01)00858254006015(17)241028(10)66793710 (01)00858254006015(17)241103(10)66897727 (01)00858254006015(17)241113(10)66947733 (01)00858254006015(17)241117(10)67054619 (01)00858254006015(17)241130(10)67308611 (01)00858254006015(17)250322(10)69548248 (01)00858254006015(17)250419(10)69548249 (01)00858254006015(17)250423(10)69548250 (01)00858254006015(17)250427(10)69567357 (01)00858254006015(17)250511(10)69567358 (01)00858254006015(17)250515(10)70322501 (01)00858254006015(17)250607(10)70523716 (01)00858254006015(17)250523(10)70597994 (01)00858254006015(17)250525(10)70751594 (01)00858254006015(17)250606(10)70791660 (01)00858254006015(17)250613(10)71054985 (01)00858254006015(17)250620(10)71197454 (01)00858254006015(17)250703(10)71197455 (01)00858254006015(17)250801(10)71197574 (01)00858254006015(17)250622(10)71319092 (01)00858254006015(17)250630(10)71361455 (01)00858254006015(17)250712(10)71589682 (01)00858254006015(17)250721(10)71780480 (01)00858254006015(17)250724(10)71780481 (01)00858254006015(17)250821(10)71974718 (01)00858254006015(17)250803(10)72167973 (01)00858254006015(17)250821(10)72216631 (01)00858254006015(17)250809(10)72216758 (01)00858254006015(17)250810(10)72282293 (01)00858254006015(17)

UPC codes

00858254006015, 00858254006008, 00858254006244, 00858254006251

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify