Instrumentation Laboratory: GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Glucose and Lactate, Part Numbers 00024315009, 00024330009, 00024345009, 00024360009

Hazard

Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.

Details from the source

Reason for recall: Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance. Distribution: US States: MA, MN, NY. India. Quantity: 2448 units Lot/code info: Part No. 00024315009: UDI-DI: 08426950082491, Lot Numbers: 323442F, 323142D, 323042K, 323142A, 323342K, 326142H, 323742C, 325442D, 325442E, 323042E, 324342A, 323042G, 323342H, 323542E, 324542A, 323542A, 324342B, 324042F, 323442H, 326442C, 325442F, 325942E, 326342B, 326142F, 325842C, 326442F; Part No. 00024330009: UDI-DI: 08426950082521, Lot Numbers: 323152A, 323152B, 324252B, 323552I, 323652I, 326152H, 323752B, 323352G, 323352H, 323052G, 323552A, 324352B, 324552A, 324052F, 323052E, 323352I, 323552F, 323352K, 325452F, 326352BC; Part No. 00024345009: UDI-DI: 08426950082552, Lot Numbers: 323462F, 323062K, 323162A, 323162C, 323562F, 323562I, 323662I, 323762D, 323762G, 323762J, 323862A, 323462C, 325762B, 323662D, 324462C, 324562A, 323362F, 325462F, 326362B; Part No. 00024360009: UDI-DI: 08426950082576, Lot Numbers: 323072J, 323072K, 323672I, 323772J, 324172D, 324372B, 323572A, 324072F, 325472F;

UPC codes

00024315009, 08426950082491, 00024330009, 08426950082521, 00024345009, 08426950082552, 00024360009, 08426950082576

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify