Hazard

Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.

Details from the source

Reason for recall: Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance. Distribution: US States: MA, MN, NY. India. Quantity: 2448 total Lot/code info: Part No. 00024307504: UDI-DI: 08426950087922, Lot Numbers: 324230M, 323630G; Part No. 00024330004: UDI-DI: 08426950082507, Lot Numbers: 325655C, 322355C, 323755H, 323855E, 326255D; Part No. 00024345004: UDI-DI: 08426950082538, Lot Numbers: 325660C, 324160J, 323760H

UPC codes

00024307504, 08426950087922, 00024330004, 08426950082507, 00024345004, 08426950082538

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify