Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Reason for recall: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection Distribution: Worldwide - US Nationwide distribution. Quantity: N/A Lot/code info: Model Number (UDI-DI): URF-P7 (04953170403811), URF-P7R (04953170403835), URF-V2 (04953170343582), URF-V2R (04953170343612), URF-V3 (04953170403385), URF-V3R (04953170403392).
04953170403811, 04953170403835, 04953170343582, 04953170343612, 04953170403385, 04953170403392
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.