Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Reason for recall: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection Distribution: Worldwide - US Nationwide distribution. Quantity: 1577 units Lot/code info: Model Number (UDI-DI): URF-P6 (04953170340802), URF-P6R (04953170340833).
04953170340802, 04953170340833
Official record: FDA (Device enforcement) ↗