Stryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing.
Reason for recall: Stryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing. Distribution: Domestic distribution to consignees in Colorado, Tennessee, and Utah. Quantity: 3 Lot/code info: UDI-DI 00869346000200 Serial Numbers AIRO-0124, AIRO-0145, AIRO-0288