Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Reason for recall: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection Distribution: Worldwide - US Nationwide distribution. Quantity: 46 units Lot/code info: Model Number (UDI-DI): ENF-T3 (04953170307188), ENF-VT2 (04953170339509), ENF-VT3 (04953170411526).
04953170307188, 04953170339509, 04953170411526
Official record: FDA (Device enforcement) ↗