← Search recalls
FDA (Device enforcement)Recalled 2023-09-25class ii

Aizu Olympus Co., Ltd.: Rhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and airway anatomy (including nasopharyngeal and trachea). Model Numbers: ENF-T3, ENF-VT2, ENF-

Hazard

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Details from the source

Reason for recall: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection Distribution: Worldwide - US Nationwide distribution. Quantity: 46 units Lot/code info: Model Number (UDI-DI): ENF-T3 (04953170307188), ENF-VT2 (04953170339509), ENF-VT3 (04953170411526).

UPC codes

04953170307188, 04953170339509, 04953170411526

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.