Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Reason for recall: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection Distribution: Worldwide - US Nationwide distribution. Quantity: 3 units Lot/code info: Model Number (UDI-DI): CHF-BP30 (04953170068751), CHF-CB30L (04953170340154), CHF-P60 (Not recorded).
04953170068751, 04953170340154
Official record: FDA (Device enforcement) ↗