SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198
Check against the official notice below — it lists the full affected range.
Wissner-Bosserhoff GmbH is recalling the SafeControl handset, a component of the Sentida bed that controls bed positioning, in model numbers 01-011539 and 01-011541. The handset has a defect where it resets to Resident mode when power is disconnected and then restored, allowing the backrest, legrest, and height adjustment to be moved when they may not be intended to be. A total of 77 units were distributed nationwide in North Carolina.
The recall notice does not specify a remedy or instructions for affected consumers.
Written by software from the official notice below. Not reviewed by a lawyer.
When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved.
Reason for recall: When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved. Distribution: US Nationwide distribution in the states of NC. Quantity: 77 units Lot/code info: UDI/DI 4050621100401, 19-0990 UDI/DI 4050621100012, 19-1062 UDI/DI 4050621100432, 19-1152 UDI/DI 4050621100449, 19-1119 UDI/DI 4050621100470, 19-1155 UDI/DI 4050621100487, 19-1156