← Search recalls
FDA (Device enforcement)Recalled 2023-09-01class ii

Convergent Dental: Solea Models 2.0 and 3.0 Laser Surgical Instrument, Foot Pedal assembly SUB-00434 with foot pedal CUS-00226 Rev. K reworked

Hazard

It has been found that potential unintended laser activation can occur without foot pedal depression.

Details from the source

Reason for recall: It has been found that potential unintended laser activation can occur without foot pedal depression. Distribution: Worldwide - US Nationwide distribution in the states of AK, AL, CA, CT, FL, KY, ME, MI, MO, MT, NC, OH, OR, PA, TN, TX, UT, VA, WA, WI and the country of Canada. Quantity: 39 units Lot/code info: Solea Model 3.0: UDI-DI: 00850634007018; Serial Numbers: 0310516, 0310517, 0310518, 0310519, 0310520, 0310521, 0310522, 0310523, 0310524, 0310525, 0310526, 0310527, 0310528, 0310529, 0310530, 0310531, 0310532, 0310533, 0310534, 0310535, 0310536, 0310537, 0310538, 0310539, 0310540, 0310541, 0310542, 0310543, 0310317, 0310321, 0310322, 0300730; Solea Model 2.0: UDI-DI: 00850634007001; Serial Numbers: 0200822, 0200807, 0200810, 0200625, CDS2-U33W-1B-DE-15, 0200819, 0200401

UPC codes

00850634007018, 00850634007001

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.