Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false resistance for clinical samples if the initial Quality Control (QC) testing of the panel batch passes, delaying treatment of the patient with the most appropriate antibiotic
Reason for recall: Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false resistance for clinical samples if the initial Quality Control (QC) testing of the panel batch passes, delaying treatment of the patient with the most appropriate antibiotic Distribution: US Nationwide distribution. Quantity: 13,900 EA Lot/code info: UDI: (10) 1355046 (17) 230131 (30) 0025 (01) 30382904486076 Lot Number: 1355046 Exp. Date: 31JAN2023
30382904486076
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.