FDA (Device enforcement)Recalled 2024-09-25class i

Elekta Instrument AB: Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Kit fo

Hazard

Elekta has become aware that Disposable Biopsy Needles can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles.

Details from the source

Reason for recall: Elekta has become aware that Disposable Biopsy Needles can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles. Distribution: US: AZ, TX, FL, CA, NY, GA, VA, NJ, MO, PA, WA, Quantity: 111 kits (666 needles) Lot/code info: Batch Number 873874875R Package including 6 needles: UDI: (01)0 7340048 30034 6 (17) 260301 (10) 873874875R (240) 911933 Package including 1 needle: UDI: (01)0 7340048 30800 7 (17) 260301 (10) 873874875R (240) 911761 Inner sterile bag including 1 needle: UDI: (01)0 7340048 30801 4 (17) 260301 (10) 873874875R (240) 912465

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify