FDA (Device enforcement)Recalled 2024-07-16class i

Smiths Medical ASD Inc.: CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems: 21-2101-0100-50, 21-2

Hazard

Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pack case. If this issue occurs, the battery pack charging circuit may become inoperable.

Details from the source

Reason for recall: Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pack case. If this issue occurs, the battery pack charging circuit may become inoperable. Distribution: Worldwide Distribution Quantity: 138039 units Lot/code info: All serial numbers.

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.