Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
Reason for recall: Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement. Distribution: US: OUS: Quantity: 22 units Lot/code info: UDI:(01)06971576831036/ Serial Numbers: 353020 353021 353023 353045 353049 354001 354002 354003 354004 354005 354006 354007 354008 354009 354010 354011 354012 354013 354014 354015 354016 354017
06971576831036
Official record: FDA (Device enforcement) ↗