After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay.
Reason for recall: After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay. Distribution: US: MT, MO, NC, SC, MN, WA, KY, TN, AR, IL, IA, MI, WV, KS, CA, HA, CO, LA, OK, ME, KY, NY, PA, GA, FL Quantity: 65 Lot/code info: Model/UDI-DI, with software: Aplio i900 TUS-AI900/ 04987670102523, Aplio i800 TUS-AI800/ 04987670102516, Aplio i700 TUS-AI700/ 04987670102509, With System Version:V4.6/V5.0/V5.1, SP0005 to SP1014, or V6.0/V6.5 SP0000 to SP1009; Aplio a550 CUS-AA550/ 04987670103353, Aplio a450 CUS-AA450/ 04987670103346, With System Version: V3.0, SP0005 to SP1014, or V4.0/V4.5, SP0000 to SP1009
04987670102523, 04987670102516, 04987670102509, 04987670103353, 04987670103346