There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
Reason for recall: There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope. Distribution: US Nationwide distribution Quantity: 32211 devices Lot/code info: Model BF-1TH190 UDI-DI: 4953170335181; Model BF-H190 UDI-DI: 4953170335174; Model BF-Q170 UDI-DI: 4953170342912 Model BF-Q190 UDI-DI: 4953170335198